MICROLIFE AG1-40 инструкция по эксплуатации онлайн - страница 12


Инструкция MICROLIFE AG1-40 для устройства тонометр содержит страницы на русском языке.

Размер файла: 315.23 kB. Состоит из 64 стр.


Вы можете скачать pdf файл этой инструкции: Скачать PDF


Предыдущая
Следующая
background image

9

4.4. Manipulate the push-button valve

The special push-button valve of BP AG1-40 allow user to deflate the manometer gradually
by pushing the button of air release valve. Turn the screw to tight up the valve before inflating.
After the pressure is high enough, deflate the manometer by slightly press the button so that
the pressure reduces at 2 - 3 mmHg per second suitable for taking readings. Press the button
repeatedly until the deflation is completed. Press the button more hardly, the air goes out more
rapidly. Holding the button will defalte the manometer continously.

4.5. Measuring procedure 

4.5.1. Put the chestpiece under the cuff 

 

The chestpiece shall not be placed ON or INTO the cuff, it 
shall be placed either under the cuff, or 1 - 2 cm below it. The 
chestpiece is then placed correctly , when the Korrotkoff’s 
sound appears strongest («loudest»).
Make sure the chestpiece is in contact with skin and above 
the brachial artery. Wear the binaural (earpieces) properly to 
check the Korotkoff’s sound during measurement.
Before using the stethoscope, be sure there is no crack on 
the diaphragm, earpieces, and tubing. Any improper setup 
or damage of the stethoscope will cause distorted sound or 
poor sound transmission to make inaccurate reading.

4.5.2. Inflating the cuff

Close the air valve by turning the screw clockwise. Do not 
over-tighten. Squeeze the inflation bulb with the hand at 
a steady rate until the pointer on the gauge is 30 mmHg 
above your normal systolic pressure value. If you are not 
sure the value, inflate to 200 mmHg first.


Что удобнее для чтения книг?

Планшет
Электронная книга
Смартфон
Книга в бумажном переплёте
Не читаю книг

QR код manualbase.ru
©  2010- ManualBase.ru