5

BP A90

EN

8. Safety, Care, Accuracy Test and Disposal

Safety and protection

This instrument may only be used for the purposes described in 

this booklet. The manufacturer cannot be held liable for damage 

caused by incorrect application.

This instrument comprises sensitive components and must be 

treated with caution. Observe the storage and operating condi-

tions described in the «Technical Specifications» section!

Protect it from:

- water and moisture

- extreme temperatures

- impact and dropping

- contamination and dust

- direct sunlight

- heat and cold

The cuffs are sensitive and must be handled with care.

Only pump up the cuff once fitted.

Do not use the instrument close to strong electromagnetic fields 

such as mobile telephones or radio installations.

Do not use the instrument if you think it is damaged or notice 

anything unusual.

Never open the instrument.

If the instrument is not going to be used for a prolonged period 

the batteries should be removed.

Read the additional safety instructions in the individual sections 

of this booklet.

Ensure that children do not use the instrument unsuper-

vised; some parts are small enough to be swallowed.

Instrument care

Clean the instrument only with a soft, dry cloth.

Accuracy test

We recommend this instrument is tested for accuracy every 

2 years or after mechanical impact (e.g. being dropped). Please 

contact Microlife-Service to arrange the test (see foreword).

Disposal

Batteries and electronic instruments must be disposed of in 

accordance with the locally applicable regulations, not with 

domestic waste. 

9. Guarantee

This instrument is covered by a 

3 year guarantee

 from the date of 

purchase. The guarantee is valid only on presentation of the guar-

antee card completed by the dealer (see back) confirming date of 

purchase or the receipt.

Batteries, cuff and wearing parts are not included.

Opening or altering the instrument invalidates the guarantee.

The guarantee does not cover damage caused by improper 

handling, discharged batteries, accidents or non-compliance 

with the operating instructions.

Please contact Microlife-Service (see foreword).

10. Technical Specifications

This device complies with the requirements of the Medical Device 

Directive 93/42/EEC.

Technical alterations reserved.

Operating temperature:

10 - 40 °C / 50 - 104 °F

15 - 95 % relative maximum humidity

Storage temperature:

-20 - +55 °C / -4 - +131 °F

15 - 95 % relative maximum humidity

Weight:

465 g (including batteries)

Dimensions:

146 x 47 x 64 mm

Measuring procedure:

oscillometric, corresponding to Korot-

koff method: Phase I systolic, Phase V 

diastolic

Measurement range:

20 - 280 mmHg – blood pressure

40 - 200 beats per minute – pulse

Cuff pressure display 

range:

0 - 299 mmHg

Resolution:

1 mmHg

Static accuracy:

pressure within ± 3 mmHg

Pulse accuracy:

± 5 % of the readout value

Voltage source:

4 x 1.5 V Batteries; size AA

Reference to 

standards:

EN 1060-1 /-3 /-4; IEC 60601-1;

IEC 60601-1-2 (EMC)