Important Safety Information
Introduction
1. Overview
Main unit:
Display:
2.1 Installing/Replacing the Batteries
2. Preparation
1.
2.
3.
1.
1.
2.
2.
2.2 Setting the Date and Time
2.3 Setting the Advanced Positioning Sensor
3. Using the Unit
1.
3.1 Applying the Wrist Cuff
3.3 Taking a Reading
3.2 How to Sit Correctly
Setting the Wrist for Measurement
Setting the Buzzer
Resetting to the Default Setting
Relax your wrist and hand.
3.4 Using the Memory Function
Do not bend
your wrist back.
Do not clench
your fist.
Do not bend
your wrist forward.
Digital Automatic
Wrist Blood Pressure
Monitor
Model R5 Prestige
Instruction Manual
Thank you for purchasing the OMRON R5 Prestige Digital Automatic Wrist
Blood Pressure Monitor.
The OMRON R5 Prestige is a compact and easy to use blood pressure
monitor, operating on the oscillometric principle. This well designed wrist
monitor comes with memory for two users and a easy to read Blood pressure
level indicator allowing you to easily see if your readings are with in the
Guidelines set by the World Health Organization. It measures your blood
pressure and pulse rate simply and quickly. For comfortable controlled inflation
without the need of pressure pre-setting or re-inflation the devices uses its
advanced “IntelliSense” technology.
Intended Use
This product is designed to measure the blood pressure and pulse rate of
people within the range of the designated wrist cuff, following the instructions in
this instruction manual.
It is mainly designed for general household use. Please read the Important
Safety Information in this instruction manual before using the unit.
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure, CONSULT YOUR
DOCTOR.
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
or arteriosclerosis.
Please read this section carefully before using the unit.
Warning:
• Indicates a potentially hazardous situation which, if not avoided, could result
in death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a
doctor before using the unit, as cuff inflation can cause internal bleeding.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Consult a doctor immediately.
Caution:
• Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury to the user or patient or damage to the equipment or
other property.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express
their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or wrist cuff.
• Do not inflate the wrist cuff over 299 mmHg.
• Do not use a mobile phone or other devices that emit electromagnetic fields,
near the unit. This may result in incorrect operation of the unit.
• Do not operate unit in a moving vehicle (car, airplane).
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with
plenty of clean water.
• Use only two “AAA” alkaline (LR03) batteries with this unit. Do not use other
types of batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace both batteries at
the same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit.
• Do not take measurements after bathing, drinking alcohol, smoking,
exercising or eating.
• Do not inflate the wrist cuff when it is not wrapped around your wrist.
• Do not wash the wrist cuff or immerse it in water.
• Read and follow the “Important information regarding Electro Magnetic
Compatibility (EMC)” in the Technical Data Section.
• Read and follow the “Correct Disposal of This Product” in the Technical Data
Section when disposing of the device and any used accessories or optional parts.
A. Display
B. Battery compartment cover
C. Wrist cuff
D. MEM (Memory) button
E. O/I START ( ) button
F. User ID selection switch
G. SET button
H. Date/Time setting ( ) button
I. Date/Time setting ( ) button
J. Deflation symbol
K. Systolic blood pressure
L. Positioning indicator
M. Diastolic blood pressure
N. Memory symbol
O. Average value symbol
P. Irregular heartbeat symbol
Q. Movement error symbol
R. User ID symbol (A or B)
S. Wrist (left/right) symbol
T. Battery low symbol
U. Pulse display
V. Heartbeat symbol
W. Date/Time display
X. Buzzer symbol
Y. Blood pressure level indicator
Remove the battery cover.
1) Push the hook on the bottom of
the battery cover.
2) Pull the cover off the main unit.
Insert two 1.5V “AAA” alkaline (LR03)
batteries as indicated in the battery
compartment and set the date and
time (refer to section 2.2).
Replace the battery cover.
The Advanced Positioning Sensor will assure you the unit is in the correct
position level with the user’s heart before the monitor starts to take the
measurement.
Selecting the User ID
Before setting the Advanced Positioning Sensor, select your User ID.
Setting the Positioning Indicator
Notes:
• If the battery low symbol ( ) appears on the display, turn the unit off
then replace both batteries at the same time.
• The measurement values continue to be stored in memory even after
the batteries are replaced.
Disposal of used batteries should be carried out in accordance with the
national regulations for the disposal of batteries.
Set the monitor to the correct date and time before
taking a measurement for the first time.
Press the button.
Do not apply over clothing.
Place the wrist cuff over your wrist.
Wrap the wrist cuff around your wrist.
Your thumb should face upward.
Make sure that the wrist cuff does not cover
the protruding part of the wrist bone (ulna) on
the outside of the wrist.
Notes:
• You can take a measurement on either your
left or right wrist. Refer to 2.3 about how to
modify the settings.
• The blood pressure can differ between the right
arm and the left arm, and therefore also the
measured blood pressure values can be
different. OMRON recommends to always use
the same arm for measurement. If the values
between the two arms differ substantially,
please check with your physician which arm to
use for your measurement.
• Wrap the wrist cuff securely around the wrist
for taking accurate measurements.
Press the O/I START button to turn the monitor off.
Notes:
• To reset the date and time, hold down the button while the power
is off.
• If the batteries have been removed for 30 seconds or more, the date
and time setting will need to be reset.
1) Push the button to change.
• Hold down to advance the
digit rapidly.
2) Push the button to confirm.
Next setting appears.
3.
Push the button to choose which mode of Positoning
indicator you wish to have.
While setting the Positioning indicator the blue light flashes.
on1
: After the Positioning indicator lights in blue for more than 2
seconds, or in orange for more than 5 seconds, measurement
starts automatically, even though the monitor is not in proper
position.
on2
: The Positioning indicator lights in blue for more than 2 seconds,
measurement starts automatically. Measurement will not start if it
lights in orange.
oFF
: Measurement starts without the Positioning indicator.
Push the button to confirm your choice and move
onto the next step.
Note: If you choose “oFF” for the Positioning indicator, the monitor will
skip the process of choosing which wrist.
Year
Month
Day
Hour
Minute
To take a measurement, you need to be relaxed and comfortably seated, under
comfortable room temperature. No eating, smoking or exercising 30 minutes
before taking a measurement.
• Sit on a chair with your feet flat on the floor.
To Delete All the Values Stored in Memory
The values stored in the memory are deleted by user ID.
You cannot partially delete values stored in the memory. All values for the
user you select will be deleted.
When the memory symbol ( ) appears, first press the MEM button. Then
while holding it down, press the O/I START button simultaneously for about
2-3 seconds.
Note: You cannot partially delete the stored readings.
Sit upright with your
back straight.
The cuff should be at the
same level as your heart.
First
Second
Do not use your other hand
to support the wrist cuff.
This can result in inaccurate
measurement results.
Notes:
• This section is the example for the measurement with the default
setting. To change the setting refer to section 2.3.
• To cancel a measurement, press the O/I START button at any time
during measurement.
• Remain still while taking a measurement.
Select your User ID.
Always use the same user ID when taking a measurement.
The unit stores the measurement values in the selected user ID
memory.
Notes:
• If “GUEST” is selected, A or B will not be displayed.
• If “GUEST” is selected, the measurement results are not stored in the
memory.
2.
Press the O/I START button.
While your position is improper, the Positioning indicator lights in
orange, when you are in proper position the buzzer beeps “beep beep
beep” and the Positioning indicator turns to blue.
The monitor automatically stores the result up to 90 sets each for both
user A and B. It can also calculate an average reading based on the
measurements from the last three readings taken within 10 minutes. If
there are only two readings in memory for that period, the average will
be based on two readings. If there is one reading in memory for that
period, the average will be based on one reading.
Note: If the memory is full, the monitor will delete the oldest readings.
Important:
• Recent research suggests that the following
values can be used as a guide to high blood
pressure for measurements taken at home.
Systolic Blood Pressure Above 135 mmHg
Diastolic Blood Pressure Above 85 mmHg
This criteria is for home blood pressure
measurement.
• Your blood pressure monitor includes an irregular
heartbeat feature. Irregular heartbeats can influence
the results of the measurement. The irregular
heartbeat algorithm automatically determines if the
measurement is usable or needs to be repeated. If
the measurement results are affected by irregular
heartbeats but the result is valid, the result is shown
together with the irregular heartbeat symbol ( ). If
the irregular heartbeats cause the measurement to
be invalid, no result is shown. If the irregular
heartbeat symbol ( ) is shown after you have taken a measurement,
repeat the measurement. If the irregular heartbeat symbol ( ) is
shown frequently, please make your doctor aware of it.
3.
4.
Undo the wrist cuff and remove the unit.
Press the O/I START button to turn the monitor off.
The monitor automatically stores the measurement in its memory.
It will automatically turn off after 2 minutes.
Average Value
Previous Reading
Alternating
date/time display
View the Average
Value
View Previous Reading
View Readings Stored
in Memory
• Hold down to view
readings rapidly.
4.3 Maintenance
To protect your unit from damage, please avoid the following:
• Subjecting your unit to extreme temperatures, humidity,
or direct sunlight.
• Washing the cuff or exposing the cuff or unit to water.
• Disassembling the unit.
• Subjecting the unit to strong shocks or vibrations.
Dropping the Unit.
• Cleaning the unit with volatile liquids.
•
The unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and soap to clean the cuff.
• Keep the unit in its storage case when not in use.
Do not store the unit in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity,
direct sunlight, dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will
be at risk of falling.
Calibration and Service
• The accuracy of this blood pressure monitor has been
carefully tested and is designed for a long service life.
• It is generally recommended to have the unit inspected
every two years to ensure correct functioning and
accuracy. Please consult your authorised OMRON
dealer or the OMRON Customer Service at the address
given on the packaging or attached literature.
5. Technical Data
Note: Subject to technical modification without prior notice.
Product Description
Model
Display
Measurement Method
Measurement Range
Accuracy
Inflation
Deflation
Memory
Power Source
Battery Life
Applied Part
Protection Against
Electric Shock
Operating temperature/
Humidity
Storage temperature/
Humidity/ Air pressure
Console Weight
Outer Dimensions
Measurable
circumference of wrist
Cuff Material
Package Content
Digital Automatic Wrist Blood Pressure Monitor
OMRON R5 Prestige (HEM-6052-RU)
LCD Digital Display
Oscillometric method
Pressure: 0 mmHg to 299 mmHg
Pulse: 40 to 180 beats/min.
Pressure: ±3 mmHg
Pulse: ±5% of reading
Automatic inflation by pump
Automatic rapid deflation
90 Measurements with date and time for each user
(A and B)
2 “AAA” alkaline (LR03) batteries 1.5V
Approx. 300 measurements with new alkaline
batteries at a room temperature of 23˚C
Internally powered ME equipment
+10˚C to +40˚C / Maximum: 30 to 85% RH
-20˚C to +60˚C / Maximum: 10 to 95% RH /
700-1060 hPa
Approximately 114g without batteries
Approximately 70 (w) mm × 70 (h) mm × 21 (d) mm
without the wrist cuff
Approximately 13.5 to 21.5 cm
Nylon and polyester
Main unit, storage case, battery set, instruction
manual, guarantee card, blood pressure pass
= Type B
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
• This blood pressure monitor is designed according to the European Standard
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements
and Part 3: Supplementary requirements for electromechanical blood
pressure measuring systems.
• This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co. Ltd., Japan. The Core component for OMRON blood
pressure monitors, which is the Pressure Sensor, is produced in Japan.
This marking shown on the product or its literature, indicates that it
should not be disposed of, with other household wastes at the end
of its working life. To prevent possible harm to the environment or
human health from uncontrolled waste disposal, please separate
this from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they can
take this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions
of the purchase contract. This product should not be mixed with other commercial
wastes for disposal.
This product does not contain any hazardous substances.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
IM-HEM-6052-RU-05-10/2011
5329896-0D
ulna
1.
2.
Push the button.
Push the button to confirm your choice and move
onto the next step.
2.
2.
3.
1.
1.
Push the button to toggle between “ ” (Left)
and “ ” (Right).
: Taking a reading on the left wrist.
: Taking a reading on the right wrist.
Push the button to choose which mode of the buzzer
you wish to have.
If the Positioning indicator is “on1” or “on2”:
on1
: The buzzer sounds only when the monitor is in proper position.
on2
: The buzzer sounds a series of two short blips if your wrist is too
far away from the measuring position, and sounds slowly when
your wrist is in proper position.
oFF
: The buzzer will not sound.
If the Positioning indicator is “oFF”:
on
: The buzzer sounds when pressing any buttons.
oFF
: The buzzer will not sound.
Push the button to confirm.
Blue light will flash.
Press the O/I START button to save settings and turn the
monitor off.
The positioning indicator is set “on1”, the wrist for measurement is set “ ”,
and the buzzer is set “on1” as default.
To reset to the default setting, press and hold down the SET button while in the
setting mode, then press the O/I START button simultaneously for more than 2
seconds.
User A
User B
The orange light
The wrist cuff will start to inflate automatically.
The blue light
Lower than
the heart.
Higher than
the heart.
Far away
from the
heart.
Keep at the
center of
the chest.
START
INFLATE
DEFLATE
COMPLETE
Alternating
date/time display
4. Troubleshooting and Maintenance
4.1 The Icons and Error Messages
Error Display
Cause
Remedy
Irregular or weak pulses
are detected.
Movement during
measurement.
The batteries are low.
The batteries are
exhausted.
The posture has
changed during
measurement.
Remove the wrist cuff. Wait
2-3 minutes and then take
another measurement.
Repeat the steps in section
3.3. If this error continues to
appear, contact your doctor.
Carefully read and repeat the
steps in section 3.3.
You should replace them with
new ones ahead of time.
Refer to section 2.1.
You should replace them with
new ones at once.
Refer to section 2.1.
Carefully read and repeat the
steps listed under section 3.3.
Cuff is over inflated.
The wrist cuff is not
wrapped securely.
Carefully read and repeat the
steps listed under section 3.1.
Do not change the user ID
during measurement.
The user ID has
changed during
measurement.
Carefully read and repeat the
steps listed under section 3.3.
An Er mark with a
code/number indicates
the device has a
hardware failure.
Consult your OMRON
retail outlet or distributor.
Blink
Lit
User A
User B
4.2 Troubleshooting
Problem
Cause
Remedy
The reading is
extremely low (or
high).
The wrist cuff is not at
heart level.
The cuff is not
wrapped snugly
around the wrist.
Your arms and
shoulders are tense.
Movement or talking
during measurement.
Measure while in the
correct posture. Refer
to section 3.2.
Wrap the cuff correctly.
Refer to section 3.1.
Relax and try taking the
measurement again.
Refer to section 3.3.
Wrist cuff pressure
does not rise.
Air is leaking from the
wrist cuff.
Consult your OMRON
retail outlet or
distributor.
Wrist cuff deflates too
soon.
The wrist cuff is
loose.
Apply the cuff correctly
so that it is securely
wrapped around the
wrist. Refer to section
3.1.
Remain still and do not
talk during
measurement. Refer to
section 3.3.
Problem
Cause
Remedy
The unit loses power
during measurement.
Nothing happens
when you press the
buttons.
The batteries are
exhausted.
Replace the batteries
with new ones.
The batteries are
exhausted.
Replace the batteries
with new ones.
The batteries have
been inserted
incorrectly.
Insert the batteries with
the correct (+/-) polarity.
Other problems.
• Press the O/I START button and repeat
measurement.
• If the problem continues, try replacing the
batteries with new ones.
If this still does not solve the problem, contact
your OMRON retail outlet or distributor.
The blood pressure is different each time.
The reading is extremely low (or high).
Blood pressure readings
constantly vary with time
of day and how relaxed
you are. Take several
deep breaths and try to
remain relaxed before
taking a measurement.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile
(cellular) telephones, medical devices in use may be susceptible to
electromagnetic interference from other devices. Electromagnetic interference
may result in incorrect operation of the medical device and create a potentially
unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)
with the aim to prevent unsafe product situations, the EN60601-1-2:2007
standard has been implemented. This standard defines the levels of immunity to
electromagnetic interferences as well as maximum levels of electromagnetic
emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This may
result in incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct operation
of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at OMRON
HEALTHCARE EUROPE at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.
Made in China
Manufacturer
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp
THE NETHERLANDS
www.omron-healthcare.com
ComplectService Ltd.
13-14 building, 26 B. Tishinskiy per.
Moscow 123557, RUSSIA
www.csmedica.ru
OMRON (DALIAN) CO., LTD.
Dalian, CHINA
EU-representative
Production Facility
Exclusive
distributor in
Russia & importer
12K2607