Wrist Blood
Pressure Monitor
Model
R1
Instruction Manual
Thank you for purchasing the OMRON R1 Wrist Blood Pressure Monitor.
This remarkable, compact and easy to use instrument is ideal for people who
frequently monitor their own blood pressure. The small, pre-formed wrist cuff is
very convenient and easy to apply.
The OMRON R1 is a compact and easy to use blood pressure monitor,
operating on the oscillometric principle. It measures your blood pressure and
pulse rate simply and quickly. For comfortable controlled inflation without the
need of pressure pre-setting or re-inflation the devices uses its advanced
“IntelliSense” technology.
The unit also stores the last measurement value in memory.
Intended Use
This product is designed to measure the blood pressure and pulse rate of
people within the range of the designated wrist cuff, following the instructions in
this instruction manual.
It is mainly designed for general household use. Please read the Important
Safety Information in this instruction manual before using the unit.
Please read this instruction manual thoroughly before using the unit. For
specific information about your own blood pressure, CONSULT YOUR
DOCTOR.
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
or arteriosclerosis.
Please read this section carefully before using the unit.
Warning:
• Indicates a potentially hazardous situation which, if not avoided, could result
in death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a
doctor before using the unit, as cuff inflation can cause internal bleeding.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Consult a doctor immediately.
Caution:
• Indicates a potentially hazardous situation which, if not avoided, may result in minor or
moderate injury to the user or patient or damage to the equipment or other property.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or wrist cuff.
• Do not inflate the wrist cuff over 299 mmHg.
• Do not use a mobile phone or other devices that emit electromagnetic fields,
near the unit. This may result in incorrect operation of the unit.
• Do not operate unit in a moving vehicle (car, airplane).
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with
plenty of clean water.
• Use only two “AAA” alkaline batteries with this unit. Do not use other types of batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace both batteries at
the same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit.
• Do not take measurements after bathing, drinking alcohol, smoking,
exercising or eating.
• Do not inflate the wrist cuff when it is not wrapped around your wrist.
• Do not wash the wrist cuff or immerse it in water.
• Read and follow the “Important information regarding Electro Magnetic
Compatibility (EMC)” in the Technical Data Section.
• Read and follow the “Correct Disposal of This Product” in the Technical Data
Section when disposing of the device and any used accessories or optional parts.
Important Safety Information
Note: Subject to technical modification without prior notice.
• This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co. Ltd., Japan. The Core component for OMRON blood
pressure monitors, which is the Pressure Sensor, is produced in Japan.
• Disposal of this product and used batteries should be carried out in accordance
with the national regulations for the disposal of electronic products.
This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
This blood pressure monitor is designed according to the European Standard
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements
and Part 3: Supplementary requirements for electromechanical blood pressure
measuring systems.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile
(cellular) telephones, medical devices in use may be susceptible to
electromagnetic interference from other devices. Electromagnetic
interference may result in incorrect operation of the medical device and
create a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic
Compatibility) with the aim to prevent unsafe product situations, the
EN60601-1-2:2007 standard has been implemented. This standard defines
the levels of immunity to electromagnetic interferences as well as maximum
levels of electromagnetic emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to
this EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This
may result in incorrect operation of the unit and create a potentially unsafe
situation. Recommendation is to keep a minimum distance of 7 m. Verify
correct operation of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at
OMRON HEALTHCARE EUROPE at the address mentioned in this instruction
manual.
Documentation is also available at www.omron-healthcare.com.
This marking shown on the product or its literature, indicates
that it should not be disposed of, with other household wastes
at the end of its working life. To prevent possible harm to the
environment or human health from uncontrolled waste
disposal, please separate this from other types of wastes and
recycle it responsibly to promote the sustainable reuse of
material resources.
Household users should contact either the retailer where they purchased
this product, or their local government office, for details of where and how
they can take this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and
conditions of the purchase contract. This product should not be mixed with
other commercial wastes for disposal.
This product does not contain any hazardous substances.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
Made in China
Manufacturer
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp
THE NETHERLANDS
www.omron-healthcare.com
ComplectService Ltd.
13-14 building, 26 B. Tishinskiy per.
Moscow 123557, RUSSIA
www.csmedica.ru
OMRON (DALIAN) CO., LTD.
Dalian, CHINA
EU-representative
Production Facility
Exclusive
distributor in
Russia & importer
= Type B
Introduction
1. Overview
Main unit:
A. Display
B. Battery compartment cover
C. Wrist cuff
D. O/I START button
Display:
E
.
Systolic blood pressure
F
.
Diastolic blood pressure
G
.
Memory symbol
(Displayed
when
viewing
last value stored in memory)
H
. Battery low symbol
I
.
Heartbeat symbol
(Flashes during measurement)
J
.
Pulse
K
.
Deflation symbol
2.1 Installing/Replacing the Batteries
2. Preparation
1.
2.
3.
Put the battery cover back in place.
Note: The measurement values continue to be stored in memory even
after the batteries are replaced.
2.2 Battery Life & Replacement
- When the battery low symbol ( ) starts to blink, you will still be able to use
the unit for a short while. You should replace the batteries with new ones
as soon as possible.
- When the symbol ( ) remains lit, the batteries are exhausted. You should
replace the batteries with new ones at once.
If the battery low symbol ( ) appears on the display,
replace both batteries at the same time.
Remove the battery compartment cover
by pulling it off in the direction of the
arrow.
1) Push the hook on the bottom of the
battery cover.
2) Pull the cover off the main unit.
Insert two identical 1.5V “AAA”
alkaline (LR03) batteries in the
battery compartment.
Make sure their polarity (+/-) is
aligned with the polarity (+/-) as
indicated in the battery compartment.
• Turn the unit off before replacing the batteries.
• Remove the batteries if the unit will not be used for three months or more.
• Dispose of batteries according to applicable local regulations.
• Two new identical 1.5V “AAA” alkaline batteries will last for approximately
300 measurements, when used to take two measurements a day.
• Since the supplied batteries are for monitoring use only, they may have a
shorter life and do not last for 300 measurements.
3. Using the Unit
4. Handling Errors and Problems
You can take a measurement on either your left or right wrist.
Measurements can also be made on
the right wrist.
Fit the unit on the right wrist as shown.
Product Description
Model
Display
Measurement
Measurement Range
Memory
Accuracy
Inflation
Deflation
Power Source
Battery Life
Operating Temperature/
Humidity
Storage Temperature/
Humidity
Weight of Main Unit
Outer Dimensions
Measurable
circumference of wrist
Cuff material
Package Content
Taking a reading on the right wrist
1.
2.
Hold the bottom part of the wrist
cuff and wrap it around the wrist
while pulling so that it fits
comfortably.
1.
Press the O/I START button to start measurement.
The wrist cuff will start to inflate automatically after the
power is turned on.
2.
After the unit has detected your blood pressure and
pulse rate, the cuff automatically deflates and your blood
pressure and pulse rate are displayed.
3.
4.
Undo the cuff and remove the unit.
5.
3.1 Applying the Wrist Cuff
This unit automatically stores the last measurement value.
1.
3.3 Taking a Reading
6. Technical Data
Correct posture during measurement helps you get accurate results.
You should also try to measure your blood pressure at the same time each day.
3.2 Correct Posture
Sit comfortably with your
back straight, hold your arm
across your chest and relax.
To Delete the Value Stored in Memory
Relax your wrist and hand.
4.1 Error Messages
Error Symbol
Cuff over inflated.
Repeat measurement.
Remain still and do not move
during measurement.
(Refer to “3.3 Taking a
Reading”.)
Movement during
measurement.
The wrist cuff
is not fastened
securely.
Carefully read and repeat
the steps listed under
“3.1 Applying the Wrist Cuff”.
An Er mark with
a code/number
indicates the
device has a
hardware failure.
Consult your OMRON retail
outlet or distributor.
The battery
power is low.
Replace the batteries with two
new “AAA” alkaline (LR03)
batteries.
Blinks or appears
continuously
Wrist cuff pressure
does not rise.
Air is leaking from
the wrist cuff.
Consult your OMRON
retail outlet or distributor.
The wrist cuff is loose.
Apply the cuff correctly
so that it is securely
wrapped around the wrist.
(Refer to “3.1 Applying
the Wrist Cuff”.)
The blood pressure is different each time.
The reading is extremely low (or high).
The unit loses power
during measurement.
Replace the batteries
with new ones.
Nothing happens
when you press
the buttons.
Replace the batteries
with new ones.
The batteries have
been inserted
incorrectly.
Insert the batteries
with the correct (+/-)
polarity.
4.2 Troubleshooting
Symptom
The reading is
extremely low (or
high).
Are you holding
the wrist cuff at
heart level?
Measure while in
the correct posture.
(Refer to “3.2 Correct
Posture”.)
Is the cuff
wrapped snugly
around the wrist?
Wrap the cuff correctly.
(Refer to “3.1 Applying
the Wrist Cuff”.)
Are your arms and
shoulders tense?
Relax and try taking
the measurement again.
(Refer to “3.3 Taking
a Reading”.)
Other problems.
3.4 Using the Memory Function
Calibration and Service
5. Maintenance and Storage
To protect your unit from damage, please avoid the following:
Movement or
talking during
measurement.
Remain still and do not
talk during measurement.
(Refer to “3.3 Taking
a Reading”.)
Wrist cuff deflates
too soon.
Cause
Correction
Cause
Correction
Notes:
• Do not take measurement after bathing, drinking alcohol, or exercising.
• Do not move or talk during measurement.
Note: Make sure that the wrist cuff does not cover the protruding part of
the wrist bone (ulna) on the outside of the wrist.
Press and hold the O/I START button until the last
measurement value is displayed along with the memory
symbol.
Do not use your other
hand to support the wrist cuff.
This can result in inaccurate
measurement results.
Do not bend
your wrist back.
Do not clench
your fist.
Do not bend
your wrist forward.
Position your arm so that
the unit is at the same
level as your heart.
Notes:
• You can take a measurement on either your left or right arm.
The blood pressure can differ between the right arm and the left arm
and therefore also the measured blood pressure values can be
different. Omron recommends to always use the same arm for
measurement. If the values between the two arms differ
substantially, please check with your physician which arm to use for
your measurement.
• To ensure correct measurement, apply the wrist cuff so that it fits
comfortably around your wrist.
• Do not apply over clothing.
Notes:
• Unless the wrist cuff is wrapped securely
around the wrist, it will not be possible to take
correct measurements.
• If you have a slender wrist, the cuff may feel
slightly loose.
However, this will not affect measurement
results and you should not force the cuff to fit.
• The remaining part of the wrist cuff can be
conveniently folded back out of the way.
• If the remaining part of the cuff is difficult to
fold back, measurement results will not be
effected.
Notes:
• Sit still and do not talk or move until the measurement is completed.
• Keep the unit at heart height until the measurement is completed.
• To stop measurement, press the O/I START button at any time during
measurement.
To View Previous Readings Stored in Memory
Systolic blood pressure
Diastolic blood pressure
Pulse
To delete stored reading, first press and hold the O/I
START button until the measurement value is displayed.
Then keep the button pressed down for about 10 seconds.
The last reading will then be deleted.
Note: After the reading has been deleted, the screen to the right of the
illustration above appears. Release the buttons to return to the
memory display.
Blood pressure readings
constantly vary with time
of day and how relaxed
you are. Take several
deep breaths and try to
remain relaxed before
taking a measurement.
The batteries are
drained.
The batteries are
drained.
• Press the O/I START button and repeat
measurement.
• If the problem continues, try replacing the
batteries with new ones.
If this still does not solve the problem, contact
your OMRON retail outlet or distributor.
• Subjecting your unit to extreme temperatures, humidity,
or direct sunlight.
• Washing the cuff or exposing the cuff or unit to water.
• Disassembling the unit.
• Subjecting the unit to strong shocks or vibrations.
Dropping the unit.
• Cleaning the unit with volatile liquids.
• The accuracy of this blood pressure monitor has been
carefully tested and is designed for a long service life.
• It is generally recommended to have the unit inspected
every two years to ensure correct functioning and
accuracy. Please consult your authorised OMRON
dealer or the OMRON Customer Service at the address
given on the packaging or attached literature.
• If the wrist cuff needs to be replaced have this done by
an authorised expert. Consult your authorised OMRON
dealer or the OMRON Customer Service.
• Do not carry out any repairs by yourself. If a defect
occurs or you have doubts about the correct functioning
of the device, consult your authorised OMRON dealer
or the OMRON Customer Service.
• The unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and soap to clean the cuff.
• Keep the unit in its storage case when not in use.
• Fold the cuff into the storage case.
Do not store the unit in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity,
direct sunlight, dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will
be at risk of falling.
Wrist Blood Pressure Monitor
OMRON R1 (HEM-6114-RU)
LCD Digital Display
Oscillometric method
Pressure: 0 mmHg to 299 mmHg
Pulse: 40 to 180 beats/min
Last measurement
Pressure: within ±3 mmHg
Pulse rate: within ±5% of reading
Automatic inflation by pump
Automatic rapid deflation
Two 1.5V “AAA” alkaline (LR03) batteries
Approximately 300 measurements when used twice
daily with alkaline batteries at a room temperature of
23°C
10°C to 40°C, 30 to 85% RH
-20°C to 60°C, 10 to 95% RH
Approximately 117 g (not including batteries)
71 mm (w) × 41 mm (h) × 70 mm (d) (not including
the wrist cuff)
Approximately 13.5 to 21.5 cm
Nylon
Main unit, storage case, two “AAA” alkaline (LR03)
batteries, instruction manual, guarantee card, blood
pressure pass
Check the measurement result.
The unit automatically stores blood
pressure and pulse rate into its
memory.
Refer to “3.4 Using the Memory
Function”.
2.
Press the O/I START button to turn
the unit off.
Note: If you forget to turn the unit off, it will
automatically shut itself off after two
minutes.
Press the O/I START button to turn off the unit.
If you forget to turn off the unit, it will shut itself off
automatically after two minutes.
ulna
Please read the instruction manual carefully before
using the device.
IM-HEM-6114-RU-05-10/2011
5329529-5C
Place the wrist cuff over your left wrist with your left thumb
facing upward.
12K2607