Thank you for purchasing the OMRON M5 Comfort Automatic Blood Pressure
Monitor.
The OMRON M5 Comfort is a compact, fully automatic blood pressure monitor,
operating on the oscillometric principle. It measures your blood pressure and
pulse rate simply and quickly. For comfortable controlled inflation without the
need of pressure pre-setting or re-inflation the device uses its advanced
“IntelliSense” technology.
Intended Use
This product is designed to measure the blood pressure and pulse rate of
people within the range of the designated arm cuff, following the instructions in
this instruction manual.
It is mainly designed for general household use. Please read the Important
Safety Information in this instruction manual before using the unit.
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure, CONSULT YOUR
DOCTOR.
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
or arteriosclerosis.
Please read this section carefully before using the unit.
Warning:
• Indicates a potentially hazardous situation which, if not avoided, could result
in death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a
doctor before using the unit, as cuff inflation can cause internal bleeding.
(AC Adapter Usage)
•
Never plug in or unplug the power cord from the electric outlet with wet
hands.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Consult a doctor immediately.
Caution:
• Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury to the user or patient or damage to the equipment or
other property.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express
their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result
in incorrect measurement results.
• Make sure that the air tube is not wrapped around other parts of your body
when taking measurements. This could result in injury when the air pressure
in the air tube is increased.
• Do not leave the cuff wrapped on the arm if taking measurements during the
night. This could result in injury.
• Do not use a mobile phone or other devices that emit electromagnetic fields,
near the unit. This may result in incorrect operation of the unit.
• Do not operate the unit in a moving vehicle (car, airplane).
• To inflate the cuff manually, refer to Section 3.3. If the cuff is over inflated, it
can cause internal bleeding.
(AC Adapter Usage)
• Use only the original AC adapter designed for this unit. Use of unsupported
adapters may damage and/or may be hazardous to the unit.
• Plug the AC adapter into the appropriate voltage outlet. Do not use a
multiple-tap.
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off
the power and unplug the power cord immediately.
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with
plenty of clean water.
• Use only four “AA” alkaline or manganese batteries with this unit. Do not use
other types of batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace all four batteries at
the same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not take measurements after bathing, drinking alcohol or caffeine,
smoking, exercising or eating.
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull on the air plug at the connection with the
main unit not the tube itself.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Read and follow the “Important information regarding Electro Magnetic
Compatibility (EMC)” in the Technical Data Section.
• Read and follow the “Correct Disposal of This Product” in the Technical Data
Section when disposing of the device and any used accessories or optional
parts.
Important Safety Information
Note: Subject to technical modification without prior notice.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile
(cellular) telephones, medical devices in use may be susceptible to
electromagnetic interference from other devices. Electromagnetic interference
may result in incorrect operation of the medical device and create a potentially
unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)
with the aim to prevent unsafe product situations, the EN60601-1-2:2007
standard has been implemented. This standard defines the levels of immunity to
electromagnetic interferences as well as maximum levels of electromagnetic
emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This may
result in incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct operation
of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at OMRON
HEALTHCARE EUROPE at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.
This marking shown on the product or its literature, indicates that it
should not be disposed of, with other household wastes at the end
of its working life. To prevent possible harm to the environment or
human health from uncontrolled waste disposal, please separate
this product from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they can
return this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions
of the purchase contract. This product should not be mixed with other commercial
wastes for disposal.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
Made in China
12E1291
Introduction
1. Overview
Main unit:
Arm cuff:
O. Date/Time display
P. User ID symbol (A/B)
Q. Systolic blood pressure
R. Diastolic blood pressure
S. Heartbeat symbol
(Flashes during measurement)
T. Battery low symbol
U. Average value symbol
V. Memory symbol
2.2 Installing/Replacing the Batteries
2.1 Using the AC Adapter
2. Preparation
1.
2.
Remove the battery cover.
2.3 Setting the Date and Time
1.
Press the button.
2.
Set the monitor to the correct date and time before
taking a measurement for the first time.
3.
Press the O/I START button to store the setting.
Notes:
• If the batteries have been removed for 30 seconds or more, the date
and time setting will need to be reset.
• If the date and time are not set, “-:--” appears during or after
measurement.
Insert four “AA” batteries
as indicated in the battery
compartment and then
replace the battery cover.
Notes:
• If the battery low symbol ( ) appears on the display, turn the unit off
then replace all batteries at the same time.
• The measurement values continue to be stored in memory even after
the batteries are replaced.
Disposal of used batteries should be carried out in accordance with the
national regulations for the disposal of batteries.
4. Troubleshooting and Maintenance
This criteria is for home blood pressure measurement.
Systolic Blood Pressure
Above 135 mmHg
Diastolic Blood Pressure
Above 85 mmHg
6. Technical Data
5. Optional Parts
4.1 The Icons and Error Messages
Blink
Lit
4.2 Troubleshooting
4.3 Maintenance
Problem
3.4 Using the Memory Function
Calibration and Service
To protect your unit from damage, please observe the following:
Cause
Remedy
To View the Average Value
To View the Readings Stored in Memory
To Delete All the Values Stored in Memory
• Do not subject the main unit and the cuff to extreme
temperatures, humidity, moisture or direct sunlight.
• Do not fold the cuff or tubing tightly.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or vibrations
(for example, dropping the unit on the floor).
• Do not use volatile liquids to clean the main unit.
• Do not wash the arm cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to clean
the arm cuff.
• Do not carry out repairs of any kind yourself. If a defect
occurs, consult your OMRON retail outlet or distributor
as mentioned on the packaging.
• The accuracy of this blood pressure monitor has been
carefully tested and is designed for a long service life.
• It is generally recommended to have the unit inspected
every two years to ensure correct functioning and
accuracy. Please consult your authorised OMRON
dealer or the OMRON Customer Service at the address
given on the packaging or attached literature.
• The unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and soap to clean the arm
cuff.
Automatic Blood Pressure Monitor
OMRON M5 Comfort (HEM-7224-ARU)
LCD Digital Display
Oscillometric method
Pressure: 0 to 299 mmHg
Pulse: 40 to 180/min.
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Fuzzy-logic controlled by electric pump
Automatic pressure release valve
90 Measurements with date and time for each user
(A and B)
DC6V 4W
4 “AA” batteries 1.5V or AC adapter
(
Adapter S-9515336-9
,
INPUT AC100-240V 50/60 Hz 0.12A)
Capacity of new alkaline batteries is approx.
900 measurements
= Type B
Internally powered ME equipment
(When using only the batteries)
= Class
II
ME equipment
(AC adapter)
+10°C to +40°C / Maximum: 30 to 85% RH
-20°C to +60°C / Maximum: 10 to 95% RH / 700-1060hPa
Approximately 380g without batteries
Approximately 210g
Approximately 123 (w) mm × 85 (h) mm × 158 (l) mm
Approximately 152 mm x 600 mm
(Cuff: arm circumference 22 to 42 cm)
Nylon and polyester
Main unit, cuff, instruction manual, quick guide,
storage case, AC adapter, battery set, guarantee
card, blood pressure pass
Storage case:
3.
Remove the arm cuff.
4.
Press the O/I START button to turn the monitor off.
The monitor automatically stores the measurement in its
memory.
It will automatically turn off after 2 minutes.
2.
Press the Memory button.
1.
Select your USER ID (A/B).
1.
Press the button, while the average value is displayed.
The monitor automatically stores the result up to 90 sets for each user
(A and B). It can also calculate an average reading based on the
measurements from the last three readings taken within 10 minutes. If
there are only two readings in memory for that period, the average will
be based on two readings. If there is one reading in memory for that
period, the average will be based on one reading.
Notes:
• If the memory is full, the monitor will delete the oldest readings.
• When viewing the reading taken without setting the date and time,
“-:--” is displayed instead of the date and time.
The Memory number appears
for a second before the pulse
rate is displayed. The newest
set is numbered “1”.
2.
Press the or button to view the readings stored in
memory.
: To the older readings
: To the more recent readings
Note: You cannot partially delete values stored in the memory.
All values for the user you select will be deleted.
Note: The irregular heartbeat symbol ( ) may also be displayed with error
messages.
The values stored in the memory are deleted by USER ID.
When the memory symbol ( ) appears, first press the Memory
button. Then while holding it down, press the O/I START button
simultaneously for about 2 - 3 seconds.
Self-diagnosis of measured results and treatment are dangerous.
Please follow the instructions of your doctor.
Warning:
• Recent research suggests that the following
values can be used as a guide to high blood
pressure for measurements taken at home.
Important:
Irregular heartbeats are
detected.
The batteries are low.
The batteries are
exhausted.
Air plug disconnected.
Cuff over inflated.
The arm cuff was inflated
above 299 mmHg
when inflating the cuff
manually.
Movement during
measurement.
Clothing is interfering with
the arm cuff.
Device error.
Remove the arm cuff. Wait
2 - 3 minutes and then take
another measurement.
Repeat the steps in section
3.3. If this error continues to
appear, contact your doctor.
Movement during
measurement.
Cuff is not applied correctly.
Carefully read and repeat
the steps in section 3.3.
Apply the arm cuff correctly.
Refer to section 3.1.
The reading is
extremely low (or
high).
Arm cuff deflates too
soon.
Arm cuff pressure
does not rise.
Cannot measure or
readings are too low
or too high.
Arm cuff not applied
correctly.
Movement or talking
during measurement.
Clothing is interfering
with the arm cuff.
The air tube is not
securely connected
into the main unit.
Air is leaking from the
arm cuff.
The arm cuff is loose.
The arm cuff has not
been inflated
sufficiently.
Apply the arm cuff
correctly.
Refer to section 3.1.
Remain still and do not
talk during
measurement.
Refer to section 3.3.
Remove any clothing
interfering with the arm
cuff.
Refer to section 3.1.
Make sure that the air
tube is connected
securely.
Refer to section 3.1.
Replace the arm cuff
with a new one.
Refer to Chapter 5.
Apply the cuff correctly
so that it is firmly
wrapped around the arm.
Refer to section 3.1.
Inflate the cuff so that it
is 30 to 40 mmHg above
your previous
measurement result.
Refer to section 3.3.
You should replace them
with new ones ahead of
time.
Refer to section 2.2.
You should replace them
with new ones at once.
Refer to section 2.2.
Insert the plug securely.
Refer to section 3.1.
Arm cuff not applied
correctly.
Apply the arm cuff correctly.
Refer to section 3.1.
Air is leaking from the arm
cuff.
Replace the cuff with the
new one.
Refer to Chapter 5.
Repeat measurement.
Remain still and do not talk
during measurement.
Refer to section 3.3.
If “E2” appears repeatedly,
inflate the cuff manually
until it is 30 to 40 mmHg
above your previous
measurement result.
Refer to section 3.3.
Do not inflate the cuff above
299 mmHg.
Refer to section 3.3.
Repeat measurement.
Remain still and do not talk
during measurement.
Refer to section 3.3.
Remove any clothing
interfering with the arm cuff.
Refer to section 3.1.
Contact your OMRON retail
outlet or distributor.
Cause
Remedy
Error Display
1.
Unplug the air tube from the air jack.
2.
Gently fold the air tube into the arm cuff.
4.4 Storage
Keep the unit in its storage case when not in use.
Note:
Do not bend the air tube excessively.
3.
Do not store the unit in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures,
humidity, direct sunlight, dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will be at an
angle.
Place the arm cuff and main unit in the
storage case.
Product Description
Model
Display
Measurement Method
Measurement Range
Accuracy
Inflation
Deflation
Memory
Rating
Power Source
Battery Life
Applied Part
Protection Against
Electric Shock
Operating
temperature/ Humidity
Storage temperature/
Humidity
Console Weight
Cuff Weight
Outer Dimensions
Cuff Dimensions
Cuff Material
Package Content
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
• This blood pressure monitor is designed according to the European Standard
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements
and Part 3: Supplementary requirements for electromechanical blood
pressure measuring systems.
• This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co. Ltd., Japan. The Core component for OMRON blood
pressure monitors, which is the Pressure Sensor, is produced in Japan
.
Problem
Cause
Remedy
• Press the O/I START button and repeat
measurement.
• If the problem continues, try replacing the
batteries with new ones.
If this still does not solve the problem, contact
your OMRON retail outlet or distributor.
Nothing happens
when you press the
buttons.
Other problems.
The batteries are
empty.
The batteries have
been inserted
incorrectly.
Replace the batteries
with new ones.
Refer to section 2.2.
Insert the batteries with
the correct (+/-) polarity.
Refer to section 2.2.
Automatic Blood
Pressure Monitor
Model M5 Comfort
Instruction Manual
A. Display
B. Memory button
C. O/I START ( ) button
D. Up/Down ( ) buttons
E. USER ID Selection Switch
(A/B)
F. Date/Time setting ( ) button
G. Air jack
H. Battery compartment
I. AC
adapter
jack
J. Blue Strip
K. Air tube
L. Air plug
M. Grip
N. Arm Cuff (Arm circumference
22-42cm)
W. Irregular heartbeat symbol
X. Movement error symbol
Y. Blood pressure level indicator
Z. Deflation symbol
AA. Cuff wrapping guide
AB. Pulse display and Memory
number
Push or buttons to
change.
Push the button to confirm.
Next setting appears.
1)
2)
Year
Month
Day
Hour
Minute
• Your blood pressure monitor includes an irregular
heartbeat feature. Irregular heartbeats can influence
the results of the measurement. The irregular
heartbeat algorithm automatically determines if the
measurement is usable or needs to be repeated. If the
measurement results are affected by irregular
heartbeats but the result is valid, the result is shown
together with the irregular heartbeat symbol ( ). If the
irregular heartbeats cause the measurement to be
invalid, no result is shown. If the irregular heartbeat
symbol ( ) is shown after you have taken a measurement, repeat
the measurement. If the irregular heartbeat symbol ( ) is shown
frequently, please make your doctor aware of it.
• If you move during measurement, the movement error
symbol ( ) will appear on the display. Keep still and
repeat the measurement.
Display:
Manufacturer
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp
THE NETHERLANDS
www.omron-healthcare.com
ComplectService Ltd.
13-14 building, 26 B. Tishinskiy per.,
Moscow 123557, RUSSIA
www.csmedica.ru
EU-representative
OMRON (DALIAN) CO., LTD.
Dalian, CHINA
Production Facility
Exclusive
distributor in
Russia & importer
IM-HEM-7224-ARU-02-06/2012
5334164-5B
Alternating
date/time
display
The bottom edge of the cuff should be 1 to 2 cm above the elbow.
Marker (arrow under tube) is centred on the middle of your inner
arm. Close the fabric fastener FIRMLY.
1.
Insert the air plug into the air jack.
2.
Hold the grip on the cuff securely
with your hand.
3.
Apply the cuff to your upper arm.
3.3 Taking a Reading
Notes:
• When you take a measurement on the right
arm, air tube will be at the side of your
elbow. Be careful not to rest your arm on
the air tube.
Notes:
• To cancel a measurement, press the O/I START button to turn the unit
off and to release the air in the arm cuff.
• Remain still while taking a measurement.
The monitor is designed to take measurements and store the
measurement value in the memory for two people using USER ID A
and USER ID B.
Select your USER ID (A/B).
3.2 How to Sit Correctly
To take a measurement, you need to be
relaxed and comfortably seated, under
comfortable room temperature. No
bathing, drinking alcohol or caffeine,
smoking, exercising or eating 30 minutes
before taking a measurement.
• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight.
• The cuff should be at the same level as your
heart.
1 to 2 cm
• The blood pressure can differ between the right arm and the left arm,
and therefore also the measured blood pressure values can be
different. OMRON recommends to always use the same arm for
measurement. If the values between the two arms differ substantially,
please check with your physician which arm to use for your
measurement.
1.
If your systolic pressure is more than
220 mmHg
After the cuff starts to inflate, press and hold the O/I START
button until the monitor inflates 30 to 40 mmHg higher than
your expected systolic pressure.
Notes:
• The monitor will not inflate above 299mmHg.
• Do not apply more pressure than necessary.
Cuff Wrapping Guide
Cuff Wrapping Guide
The Cuff Wrapping Guide is a unique feature that indicates if the
cuff is not wrapped tightly enough around the arm. Even when
the
is displayed, a blood pressure reading will be taken.
Note: This reading is
NOT
reliable due to the incorrect wrapping of the
cuff. Please wrap the cuff again, taking care to wrap it correctly
and take the measurement again. When the is displayed, the
cuff is correctly wrapped tightly enough on the arm and the
reading is accurate and reliable.
Note: Wait 2 - 3 minutes before taking another blood pressure
measurement. Waiting between readings allows the arteries to
return to the condition prior to taking the blood pressure
measurement.
Note: If there are no measurements results
stored in the memory, the screen to the
right is displayed.
3. Using the Unit
3.1 Applying the Arm Cuff
Remove tight-fitting clothing or tight rolled up sleeve from your
upper arm.
Do not place the cuff over thick clothes.
2.
Press the O/I START button.
AC Adapter S:
1.
2.
Insert the AC adapter plug
into the AC adapter jack on
the rear side of the main
unit.
Plug the AC adapter into an
electrical outlet.
To disconnect the AC adapter, unplug the AC adapter from the
electrical outlet first and then remove the AC adapter plug from the
main unit.
START
INFLATE
END
COMPLETE
DEFLATE
N
J
L
M
K
AC Adapter S
Adapter S-9515336-9
Arm Cuff
Arm circumference
22 - 42 cm
Comfort cuff-9956685-4