14
Important Notes
•
When unused for a long time or at a low temperature, the machine will take
longer to warm up time.
•
After 5 consecutive tests, allow the device to rest for 2 minutes.
•
Please store in a dry environment at room temperature.
•
Avoid extreme heat or long term sun exposure.
•
When battery indicator flashes, replace with a new battery.
•
If the difference between two successive readings is >0.1
‰
(0.01%) user
should
test again.
•
During the breath sampling, do
NOT
block the outlet.
•
An “Err” flashes on the display for 5 seconds when user fails to start the test
within 5 seconds.
Specification
Fuel cell Sensor.
Display Range: 0.00~ 4.00
‰
BAC; or 0.00~0.40% BAC or 0.00mg/L~2.0mg/L (convec-
tion factor is 2.0)
Power Input: 3 x AAA alkaline battery.
Display: Digital LCD display with bi-colour Backlighting.
Auto Power Off: 2 minutes.
Breath Sampling: about 7 seconds continuous breath sample at 20 L/min flow rate.
Dimension: 107 x 45 x 18 mm
Weight: 55 g
Operating temperature: 10
°
to + 40
°
, for optimum operation
Agency Approval: CE
Accessory: 2 pcs mouthpieces
Alert Signals
DISPLAY & ALERT
RESULT (BAC)
Green backlight and shows 0.00
‰
≤0.05
‰
BAC (0.005%BAC or 0.025 mg/l)
Alarm:
“NO DRIVING”
sigh flashes
with alarm and red backlight for 16
seconds, Show the digital reading on
the display
0.05
‰
~3.99
‰
BAC
(0.005%~0.399%BAC or 0.025~1.995mg/L)